Fingertip Oximeter

Fingertip Oximeter
Product Detailed
Special design for children Comfortable for wearing Low power and finger-out alarm Rechargable button lithium battery

 Item Number:FPX-023 

Item Name:Fingertip Oximeter

FeatureSpecial design for childrenComfortable for wearingLow power and finger-out alarmRechargable button lithium batteryAutomatic power-off when finger is not detected

Technical specifications

1,Main performance:

l  SpO2 value display

l  Pulse rate value display,bar graph display

l  Low-power indication: low-power indication symbol appears before working abnormally which is due to low-power.

l  Pulse sound indication

l  With alarm function

2,Main parameters:

Measurement of SpO2

Measuring range: 0%~100%

Accuracy: 70%~100%:±2%:Below 70% :unspecified.

Measurement of pulse rate

Measuring range:30bpm~250bpm

Accuracy: ±2 bpm or ±2% (select larger)

Resolution

SpO2 : 1%, Pulse rate: 1bpm.

Measurement Performance in Weak Filling Condition

SpO2 and pulse rate can be shown correctly when pulse-filling ratio is 0.4%. SpO2 error is ±4%, pulse rate error is ±2 bpm or ±2% (select larger).

Resistance to surrounding light

The deviation between the value measured in the condition of man-made light or indoor natural light and that of darkroom is less than ±1%.

Power supply requirement

 3.6 V DC ~ 4.2V DC.